Environment and Natural Resources Institute
|
|
 |
|
 |
Ion Chromatography
Major anion analyses are performed using IonPac As15 (4mm) column, and self regenerating suppressor ASRS Ultra (4mm); sample loop is 25 μL. Anion analyzed: | retention time (min) | | Fluoride | 3.8 | | Chloride | 4.5 | | Nitrite | 6.2 | | Sulfate | 7.2 | | Bromide | 12.6< | | Nirate | 15 | | Phosphate | 20 | Samples should be filtered through 0.45μm filters immediately after sampling and stored at 4 ºC. The shelf time of some anions is only 48 hours. The analysis is performed using calibration standard at 4 levels adjusted to concentration of the samples. Quality control and blank control is performed every 10th sample. For a full SOP click . Perchlorate is separated using IonPac AS16 (2 mm) column and self regenerating suppressor ASRS Ultra II (2mm) with external water mode. KOH | Eluent | Isocratic | 65 mM | Flow rate | <0.25 mL/min | Total time | 14 minutes | Anion analyzed: | Perchlorate | retention time | 12.5 min | Detection is performed with conductivity detector in water matrixes. For complex matrixes (e.g. tissue) the IC is connected to a mass spectrometer (Finnigan AQA) with an electrospray ionization source. The settings of mass spectrometer are: Negative mode | -ve | AQA mix | 5 V | probe temperature | 300 C | For further details on our pechlorate analyses see Dodds, et al. (2004), Anal. Bioanal. Chem. 379:881-887. © 2008 ASET Laboratory
|
|
 |
|
 |
 |
|
|
|