Text-Only Site |Campus Map |A to Z |Directory
UAA  	Logo
Environment and Natural Resources Institute
Print Friendly
BETX Analysis

 
 

This method is for peak, ceiling and Time Weighted Average (TWA) determinations of Volatile Petroleum Hydrocarbons (VPH), specifically benzene, toluene, ethylbenzene, p&m-xylene, and o-xylene.  There is the possibility that interactions between analytes may reduce the breakthrough volumes and change desorption efficiencies. Samples are analyzed using a Varian 3800 Gas Chromatograph equipped with a 2200 ion trap mass selective detector (MSD) and an 8400 auto injector. Helium is used as the carrier gas and a compatible high resolution capillary column.  Results are compared to a calibration curve.  A separate computer program is used to convert concentrations to ppbv.  This procedure assumes working knowledge of the software analysis packages currently in use. For detailed SOP click here.

 
 
UAA Logo
Page Updated: 4/17/08  By:  Susan Klein